Detecting Unit of Gamma Radiation

  • High sensitive scintillator CsI (Tl) with silicon photomultiplier
  • Availability of amplitude spectrum (1024 channels)
  • No microphonic effect
  • Meets the requirements of IEC 61000-4-2:2008, IEC 61000-4-3:2007, IEC 61000-4-4:2008 standards
  • Ingress protection rating IP67
  • An option of a waterproof housing IP68

Description

Designed to measure dose equivalent rate of gamma radiation. Used for mobile laboratories, as a part of data panels and automated radiation monitoring systems.
It is produced using a high sensitive scintillator CsI (Tl) with silicon photomultiplier and a Geiger-Muller counter. If the customer chooses so, BDBG-09S can be made of duralumin or stainless steel. BDBG-09S detecting unit can be used as a part of systems for measurement of dose equivalent rate of gamma radiation and temperature.

Purpose of Use
Measurement of ambient dose equivalent rate (DER) of gamma radiation

Features

  • High sensitive scintillator CsI (Tl) with silicon photomultiplier and Geiger-Muller counter have been used.
  • Data exchange between the detecting unit and the data display system through RS-485 interface.
  • Availability of amplitude spectrum (1024 channels).
  • Constant self-testing.
  • Availability of the statistical accuracy of measurement.
  • Ingress protection rating IP67.
  • Average service life of the detecting unit – not less than 10 years.

Detector types

Scintillation CsI (Tl) detector with a photomultiplier for measurements in the range within 0.01 μSv/h to 50 μSv/h (spectrum-dose function)
Energy-compensated Geiger-Muller counter for measurements in the range within 50 μSv/h to 1.0 Sv/h

Specifications

  • Measurement range of gamma radiation DER and main relative error
    0.05 μSv/h … 1 Sv/h
    ±(15+1/H*(10))%,where H*(10) is a numeric value of measured DER equivalent to μSv/h
  • Gamma radiation sensitivity (137Cs), not less than
    ≥ 120 cps/(µSv/h)
  • Energy range of registered gamma and X-ray radiation and energy dependence
    0.05 … 3.0 MeV
    (0.05 … 1.25; ±25 %)MeV
  • Integration time
    2 … 20 seconds
  • Setup time of operating mode of the detecting unit
    not more than 1 min
  • Operating supply voltage range of the detecting unit from external stabilized power supply
    7 … 13 V
  • Current consumption of the detecting unit for overall range of gamma radiation DER to be measured
    50 mА
  • Operating temperature range
    -30 … +50 °С
  • Relative humidity
    up to 100 % given +40 °С and lower temperatures with humidity condensation
  • Weight
    0.6 kg
  • Dimensions
    170 × 60 × 60 mm

Delivery Kit

  • BDBG-09S detecting unit of gamma radiation
  • corbel for vertical mounting of the unit
  • mounting parts kit
  • technical description and operating manual (one copy per one consignment of the detecting units)
  • logbook
  • packing box
  • Technological kit used for testing of the detecting units (one kit per one consignment of the detecting units) on demand.