Specifications
Detector | disproportionate neutron counter |
Dose rate measurement range of neutrons for Pu-α-Be, H*(10) | 0.1 μSv·h-1 ÷ 0.1 Sv·h-1 |
Dose measurement range of neutrons for Pu-α-Be, H*(10) | 0.1 μSv ÷ 1.0 Sv |
Power flux density measurement of neutrons for for Pu-α-Be
Energy range of neutrons |
1 ÷ 104s-1·сm-2
0.025 eV ÷ 14 MeV |
Energy dependence of sensitivity to neutron radiation for typical neutron spectra relative to the spectrum of Pu–Be neutron source | no more than ±40 % |
Basic measurement error | ± (25 + 5/Ах*) % |
Ingress protection | IP65 |
Overall dimensions, weight: | 295x142x100mm, 2.25kg |
* Where Ax is a non-dimensional value numerically equal to a measured value of dose rate (μSv/h) or dose (μSv).