Purpose of Use

  • Measurement of ambient dose equivalent rate (DER) of gamma and X-ray radiation
  • Measurement of ambient dose equivalent (DE) of gamma and X-ray radiation
  • Measurement of beta-particles surface flux density and beta-particles surface activity
  • Measurement of alpha-particles surface flux density and alpha-particles surface activity
  • Measurement of accumulation time of gamma and X-ray radiation DE
  • Archive of measurement results tagged to location coordinates

Specifications

Measurement of gamma and X-ray radiation parameters

  • Measurement range of gamma and X-ray DER with the BDKS-01 combined detecting unit
    0.1 µSv/h …10 Sv/h
  • Measurement range of gamma and X-ray DER with the BDKS-02 combined detecting unit
    0.1 µSv/h…2.0 Sv/h
  • Measurement range of gamma and X-ray DER with the wide range removable detector
    0.01…100 Sv/h
  • Measurement range of gamma and X-ray DE with the detector integrated in a control panel and main relative permissible error
    0.001…9 999±15 mSv%
  • Main relative permissible errors in DER measurement when calibrated with 137Cs for:
  • BDKS-01 and BDKS -02 combined detecting units

±(15+2/H*(10)) %, where H*(10) is a numeric value of measured DER in µSv/h

  • wide range removable detector
    ±(15+200/H*(10)) %, where H*(10) is a numeric value of measured DER in mSv/h
  • Energy range of measurement
    0.05…3.0MeV
  • Measurement range of DE accumulation time and measurement precision
    1 min…100 h; ±1 min per 100 h

Measurement of beta radiation parameters

  • Measurement range of surface beta-particles flux density with the BDKS-01 and BDKS -02 combined detecting units
    10…200 000 1/(cm2×min)
  • Measurement range of beta-particles surface activity with the BDKS-01 and BDKS -02 combined detecting units
    0…13500 Bq/cm2
  • Energy range of measurement with the BDKS-01 combined detecting unit
    0.3…3.0 MeV
  • Energy range of measurement with the BDKS-02 combined detecting unit
    0.15…3.0 MeV
  • Main relative permissible error in measurements of beta-particles flux density when calibrated to 90Sr+90Y
    ±(20+200/φβ) %, where φβ is a numeric value of measured surface flux density part./(cm2×min)
  • Main relative permissible error in measurements of beta-particles surface activity when calibrated to 90Sr+90Y
    ±(20+10/B) %, where B is a numeric value of measured beta-particles surface activity Bq/cm2

Measurement of alpha radiation parameters

  • Measurement range of surface alpha-particles flux density with the BDKS-02 combined detecting unit
    10…300 000 1/(cm2×min)
  • Measurement range of alpha-particles surface activity with the BDKS-02 combined detecting unit
    0…13500 Bq/cm2
  • Energy range of measurement with the BDKS-02 combined detecting unit
    from 4.0 MeV and more
  • Main relative permissible error in measurements of alpha-particles flux density when calibrated to 239Pu
    ±(15+300/φ) %, where φ is a numeric value of measured surface flux density part./(cm2×min)
  • Main relative permissible error in measurements of alpha-particles surface activity when calibrated to 239Pu
    ±(15+15/A) %, where A is a numeric value of measured alpha-particles surface activity Bq/cm2
  • Operating temperature range (digital display)
    -40…+50 °С (-40…+95)

Weight and dimensional characteristics
Weight, kg Dimensions, mm

  • BDKS-01 detecting unit without the cable 0,3 D48x165
  • BDKS-02 detecting unit without the cable 0,6 160x73x43