Purpose of Use
- Measurement of ambient dose equivalent rate (DER) of gamma and X-ray radiation
- Measurement of ambient dose equivalent (DE) of gamma and X-ray radiation
- Measurement of beta-particles surface flux density and beta-particles surface activity
- Measurement of alpha-particles surface flux density and alpha-particles surface activity
- Measurement of accumulation time of gamma and X-ray radiation DE
- Archive of measurement results tagged to location coordinates
Specifications
Measurement of gamma and X-ray radiation parameters
- Measurement range of gamma and X-ray DER with the BDKS-01 combined detecting unit
0.1 µSv/h …10 Sv/h - Measurement range of gamma and X-ray DER with the BDKS-02 combined detecting unit
0.1 µSv/h…2.0 Sv/h - Measurement range of gamma and X-ray DER with the wide range removable detector
0.01…100 Sv/h - Measurement range of gamma and X-ray DE with the detector integrated in a control panel and main relative permissible error
0.001…9 999±15 mSv% - Main relative permissible errors in DER measurement when calibrated with 137Cs for:
- BDKS-01 and BDKS -02 combined detecting units
±(15+2/H*(10)) %, where H*(10) is a numeric value of measured DER in µSv/h
- wide range removable detector
±(15+200/H*(10)) %, where H*(10) is a numeric value of measured DER in mSv/h - Energy range of measurement
0.05…3.0MeV - Measurement range of DE accumulation time and measurement precision
1 min…100 h; ±1 min per 100 h
Measurement of beta radiation parameters
- Measurement range of surface beta-particles flux density with the BDKS-01 and BDKS -02 combined detecting units
10…200 000 1/(cm2×min) - Measurement range of beta-particles surface activity with the BDKS-01 and BDKS -02 combined detecting units
0…13500 Bq/cm2 - Energy range of measurement with the BDKS-01 combined detecting unit
0.3…3.0 MeV - Energy range of measurement with the BDKS-02 combined detecting unit
0.15…3.0 MeV - Main relative permissible error in measurements of beta-particles flux density when calibrated to 90Sr+90Y
±(20+200/φβ) %, where φβ is a numeric value of measured surface flux density part./(cm2×min) - Main relative permissible error in measurements of beta-particles surface activity when calibrated to 90Sr+90Y
±(20+10/B) %, where B is a numeric value of measured beta-particles surface activity Bq/cm2
Measurement of alpha radiation parameters
- Measurement range of surface alpha-particles flux density with the BDKS-02 combined detecting unit
10…300 000 1/(cm2×min) - Measurement range of alpha-particles surface activity with the BDKS-02 combined detecting unit
0…13500 Bq/cm2 - Energy range of measurement with the BDKS-02 combined detecting unit
from 4.0 MeV and more - Main relative permissible error in measurements of alpha-particles flux density when calibrated to 239Pu
±(15+300/φ) %, where φ is a numeric value of measured surface flux density part./(cm2×min) - Main relative permissible error in measurements of alpha-particles surface activity when calibrated to 239Pu
±(15+15/A) %, where A is a numeric value of measured alpha-particles surface activity Bq/cm2 - Operating temperature range (digital display)
-40…+50 °С (-40…+95)
Weight and dimensional characteristics
Weight, kg Dimensions, mm
- BDKS-01 detecting unit without the cable 0,3 D48x165
- BDKS-02 detecting unit without the cable 0,6 160x73x43