• Continuous automatic control of density of industrial solutions, suspensions, slurries without physical contact with them;
• Control of the given density of solutions (with optional temperature sensor );
Area of application
Mining, chemical, metallurgical, construction industry, nuclear fuel cycle enterprises
Operating principle
The principle of densitometer operation is based on the transillumination of the layer of controlled liquid by the flow of gamma rays and measuring its density by degree of weakening of the flow density
Limits of density control |
from 200 to 3000 kg/м3 |
|||
The optimum measuring range of density |
100 kg/м3 |
|||
Basic (to the max. measuring range) error at P = 0,95 |
from 0,1 to 1,0 % |
|||
The least absolute error |
2,0 kg/м3 |
|||
Range of diameters of pipelines (excluding wall thickness) |
from 100 to 800 мм |
|||
Duration of measurement |
from 10 to 1000 с |
|||
Time of reaction on abrupt changing of the measured value |
from 1 to 3 с |
|||
Operation mode setup time |
No more than 30 min |
|||
Operation mode of the densimeter |
24 hours |
|||
Instability of readings within 24 hours of continuous operation |
No more than 1,0 kg/м3 |
|||
Densimeter is powered by AC of 50 ± 1 Hz, with voltage of |
from 187 to 242 В |
|||
Alternating Voltage at “dry contact ” circuit with current of up to 3A |
to 300 В |
|||
Operating temperature range: – for detection unit – for indicating device |
from – 20°С to + 50°С from 0°С to + 50°С |
|||
Relative humidity |
up to 100 % at + 35°С |
|||
Level of casing protection |
No worse than IP54 |
|||
Distance between the indication device and the detection unit |
to 1200 м |
|||
Mean time between failures |
No less than 10000 hours |
|||
Average lifetime |
No less than 10 years |
|||
Designation |
Overall dimensions |
Weight |
||
Detection unit |
Ø50×290 мм |
0,85 kg |
||
Matching unit |
180×80×60 мм |
0,6 kg |
||
Collimator |
160×160×200 мм |
11,0 kg |
||
Control block |
220×140×100 мм |
2,8 kg |
||
Process remote control |
210×120×90 мм |
1,0 kg |