Intended for:
• Detecting contamination of palms, overalls and shoes of the staff by radioactive substances
• Producing a visible signal in case if the detected level of contamination exceeds the threshold.
• Measuring surface density of beta-radiation flux
measurement range of flux density of beta particles |
10 – 1,5·104 part/ min∙cm2 |
||
Energy range of registered beta-radiation |
From 0,3 to 3,0 MeV |
||
Limit of the basic relative error of the radiometer at p = 0,95 |
± (20 + 200/Ах) % * |
||
Instability of the radiometer readings during 24 h of continuous operation |
No more than 5 % |
||
Own background of the radiometer (at MED of gamma-radiation up to 2,0 mkZv/h |
No more than 15 part/ min ∙cm2. |
||
Operation mode setup time |
No more than 1 min |
||
Working area of the detector |
160 cm2 |
||
Radiometer is resistant to the values of MED of gamma-radiation |
10 mkzv/h |
||
Radiometer is powered by AC of 50±2 Hz, voltage |
from 187 to 242 V |
||
Alternating Voltage “dry contact ” with a load of up to 1 A |
from 0 to 280 V |
||
Temperature of environment |
from -10°С to +40°С |
||
Relative humidity |
Up to 90 % at +40°С |
||
Cable length between the detection unit and the control unit |
Up to 2 m |
||
Level of casing protection |
No worse than IP54 |
||
Type of climatic performance according to GOST15150 |
UZ.1 |
||
Additional error at max. influencing factor |
No more than 10 % |
||
Mean time between failures |
No less than 10000 h |
||
Average lifetime |
No less than 10 years |
||
Designation |
Overall dimensions |
Weight |
|
Control panel |
220×140×110 mm |
3,0 kg |
|
Detection unit |
210×150×50 mm |
1,0 kg |
* – Ах –numeric measured value